关键词: Balloon wedge hepatic venous pressure catheter wedge hepatic venous pressure free hepatic venous pressure hepatic venous pressure gradient

来  源:   DOI:10.4103/0971-3026.90693

Abstract:
OBJECTIVE: To evaluate the accuracy of measurement of hepatic venous pressure gradient by catheter wedge as compared to balloon wedge (the gold standard).
METHODS: Forty-five patients having a clinical diagnosis of intrahepatic portal hypertension were subjected to the two different types of pressure measurements (catheter wedge and balloon wedge) during transjugular liver biopsy under fluoroscopic guidance.
METHODS: Spearman\'s rank correlation coefficient, Bland-Altman plot for agreement, and single measure intraclass correlation were used for analysis of data.
RESULTS: There was a close correlation between the results obtained by both the techniques, with highly significant concordance (P < 0.0001). Hepatic venous pressure gradients as measured by the catheter wedge technique were either equal to or less than those obtained by the balloon wedge technique.
CONCLUSIONS: The difference in hepatic venous pressure gradients measured by the two techniques is insignificant.